11 Localisation des pannes
Code
d'erreur
0x080251
0x080252
0x080255
0x080261
0x080271
0x080272
0x080281
0x080282
0x080291
0x080292
0x080293
236
CLAM™-2040 CL
Message d'erreur
Failed to reset the reagent
probe rotation.
Failed to reset the reagent
probe rotation.
Failed to reset the reagent
probe rotation.
Failed to rotate the reagent
probe.
Failed to reset the reagent
syringe pump.
Failed to reset the reagent
syringe pump.
The reagent probe failed to
suction.
The reagent probe failed to
discharge.
An abnormality has occurred in
the reagent probe vertical
driving motor.
An abnormality has occurred in
the reagent probe rotating
motor.
An abnormality has occurred in
the reagent syringe pump
motor.
Statut de
l'unité après
Mesure requise de l'utilisateur
erreur
STOP
• Vérifiez qu'il n'y a pas de corps étranger
dans la rainure dans laquelle la sonde du
STOP
réactif passe.
STOP
• Vérifiez que le cache des réactifs est
correctement positionné.
STOP
• Si le nombre d'utilisations indiqué au
STOP
paramètre [Syringe packing (reagent)] est
supérieur à la valeur stipulée ou a été
STOP
utilisé pendant plus de 6 mois, remplacez
l'habillage de la seringue en vous référant
au paramètre [Syringe packing (reagent)]
STOP
dans la fenêtre Pièces de rechange.
• Vérifiez que le filtre de la pompe à
seringue n'est pas obstrué.
• Si cette erreur se produit après le
remplacement de l'habillage de la
seringue (pour réactif ), vérifiez que le
piston n'est pas de travers.
STOP
• Si la sonde du réactif est obstruée, retirez
les éléments l'obstruant à l'aide du fil
fourni.
• Vérifiez que la courroie de seringue de la
pompe à seringue (côté réactif ) n'est pas
coupée.
STOP
STOP
Contactez votre représentant Shimadzu.