Code
Message d'erreur
d'erreur
Sample probe protocol error
0x0702A2
(Sample dispense)
Sample probe protocol error
0x0702A3
(Sample dispense)
Sample probe protocol error
0x0702A4
(Sample dispense)
Sample probe protocol error
0x0702B2
(Detergent suction)
Sample probe protocol error
0x0702B3
(Detergent suction)
Sample probe protocol error
0x0702B4
(Detergent suction)
Protocol command for
0x0702B6
dispensing sample with diluent
cannot finish (time over).
Protocol command for
dispensing sample with diluent
0x0702B7
cannot start (abnormal
parameter).
Protocol command for
0x0702B8
dispensing sample with diluent
cannot start (exclusive control).
Sample probe protocol error
0x0702C2
(Washing)
Sample probe protocol error
0x0702C3
(Washing)
Sample probe protocol error
0x0702C4
(Washing)
0x0702FE
Abnormal data ROM
0x0702FF
A system error occurred.
0x08**** (sonde du réactif )
Reagent probe: Faulty liquid
0x080121
level sensing
Reagent probe finds remaining
0x080123
amount of reagent is small.
Statut de
l'unité après
Mesure requise de l'utilisateur
erreur
STOP
STOP
STOP
STOP
STOP
STOP
Vérifiez que l'association du logiciel du PC
STOP
de la console et de la version du logiciel
embarqué est appropriée.
STOP
STOP
STOP
STOP
STOP
STOP
Contactez votre représentant Shimadzu.
STOP
• Placez correctement les flacons.
• Retirez les taches sur la pointe de la sonde
du réactif.
• Vérifiez qu'il n'y a pas de bulles d'air dans
le flacon positionné.
• Ajustez la sensibilité du capteur en vous
référant au paramètre [Probe (Reagent)]
dans la fenêtre Pièces de rechange.
• Placez correctement les flacons.
• Retirez les taches sur la pointe de la sonde
du réactif.
• Ajustez la sensibilité du capteur en vous
référant au paramètre [Probe (Reagent)]
dans la fenêtre Pièces de rechange.
11.2 Liste des messages d'erreur
233