DESCRIPTION
The sterilization chamber is entirely made of bright stainless steel resistant to heat and corrosion. It is
fully insulated with fibra glass to avoid the overheating of external side. The door closes sterilization
chamber hermetically by means of a silicon gum gasket resistant to heat. The sterilizer has an extractable
stainless steel tray with big holes for warm air circulation. There is a 250 Watt stainless steel armoured
electrical certified resistance characterized by high efficiency and little inertia.The automatic control
of temperature is obtained with a capillary bulb automatic thermostat for auto-adjustment up to 200°C.
If the temperature exceeds the limit 220°C, a second safety bimetallic thermostat takes control. The
working of the second thermostat causes the alarm red pilot lamp lighting.
To program the thermostat starting time, it is possible to use the timer up to 120 minutes or fixed on the
manual position.
The device is also provided with a green bright main power switch and with green, orange and red
lights for working, heating and alarm respectively. The sterilizer is delivered with certified supply
cable, one extractable stainless steel tray, two holdertray pliers.
INSTRUCTIONS
Carriage
The sterilizer is shipped in an adequate package that requires the common handling cautions for fragile
materials, as shown on the package itself. Use only the original package if the equipment must be
returned to the manufacturer.To carry the sterilizer without package, keep it in a horizontal position
and use the base as holding point.
Installation - start up
Remove the equipment from the package handling with care. Check the power supply system is correctly
grounded. Place the sterilizer on a level surface so that air can freely circulate under the equipment
bottom side. Keep the back side of the sterilizer at least 10cm. away from walls to ensure an adequate
ventilation.
Insert the safety plug of the power supply cable, firstly into the equipment socket and then into the
electric system, after being sure the voltage corresponds to the one selected on the device. Press the
green main power switch position "I". Check if the switch light is on.
Note: a typical smell of heated ceramic fibre may be present at the first power on and remain for a short
period of time.
Instructions to use
- Press the main power switch position "I".
- Turn the timer knob (nr.1) clockwise, positioning the index on the required time, up to 120 minutes.
- Check the lighting of the orange light. The orange light (nr.4) (whose lighting points out the working
of the resistance) will have some automatic clicks because of heat distribution into the sterilization
chamber, before stabilizing on the required temperature.
- When the device is cold, the required temperature will be reached in 30 minutes approx. The suggested
temperature is pointed out blue (150°-180°).
- To disconnect the timer and to make the sterilization work for an indefinite time, turn the timer knob
anticlockwise, on the "Manual" position.