12. Errors, causes, and solutions

Potential errors in the NANOCOLOR
Error
Sipper module not connect-
ed! Please check the USB
cable connection.
Please check the sipper and
the hoses.
Please check the pump head. Hose is defective.
The Sipper module does not
pump liquid.
Hardware error
The sipper module is not
detected by the spectropho-
tometer.
The spectrophotometer
shows the icon for removal
of the cuvette during the
starting sequence.
Missing response from sipper
Sipper module is connected
and activated, but cannot be
used during measurement.
In case of a persisting error, please contact MACHEREY-NAGEL or your local distributor.
38
All manuals and user guides at all-guides.com
®
FP-200 are displayed on the photometer.
Cause
Sipper is not recognized by the
spectrophotometer.
Pump is blocked.
Wrong hose was used.
Pumping head not installed
correctly.
The system has a leakage and
sucks air.
Sipper module defective.
Connectivity settings are not set
correctly.
USB cable is not connected
correctly or cable is broken.
The flow cell is still placed in the
cuvette slot of the spectropho-
tometer. The device can only be
started with an empty cuvette
slot.
Communication between sipper
module and spectrophotometer
is interrupted.
There are no sipper methods
defined.
NANOCOLOR
Solution
Cable is not installed correctly.
Wrong cable was used.
Pump head not placed correctly
on pump.
Replace the pinched hose.
Use only original hoses and
adapters.
Install pump head as described
in the user manual.
Please control all connections
for correct fixing.
Please contact
MACHEREY-NAGEL or your
local distributor.
Please check the connectivity
settings as described in section
10.1.
Please check the USB cable
connection for correct contact.
Please remove the flow cell
from the before starting the
spectrophotometer.
Please disconnect USB
connection cable and reconnect
again.
Please define one or more
sipper methods via the settings
menu of the sipper module
(section 10.3)
®