AULT FOUND
Water in the motor
• Defective O rings
No light on IMLN micro-
• 3.5 V bulb blown
motor
• Wires in cord cut
F
AULT FOUND
No air
• The air supply cord is not connected
• Insufficient air pressure
• The air supply cord is blocked
No water
• The water supply cord is not connected
• Insufficient water pressure
• The water supply cord is blocked
• Unit switched off
MICRO-MOTOR (
CONTINUED
P
OSSIBLE CAUSE
SYRINGE
P
OSSIBLE CAUSE
to the unit
13
)
R
EMEDY
• Change the O rings
• Replace the bulb
• Return to SATELEC After
Sales
R
EMEDY
• Connect the air supply
cord to the unit
• Check the air pressure and
adjust it as shown using a
pressure gauge
• Clean the air supply cord
or return to SATELEC After
Sales.
• Connect the water supply
to the unit.
• Check the water pressure
as shown using a pressure
gauge
• Clean the water supply
cord or return to SATELEC
After Service.
• Switch unit on