TM
Frontier
Centrifuge FC5707
3.6 Program
3.6.1
Storage of programs
You can store up to 99 runs with all relevant parameters, including the used rotors. You can use any free program
number and call it up again.
Put the needed rotor into the centrifuge. By pressing the key "Prog/Setup" (3) in the display "Time/Prog" (A-4)
the word "programm" (M10) appears. With the adjusting dial (4) you can chose the desired program number.
If a program number is already occupied, in the display "RPM/RCF" (A-1), the words "rotor" (M3) and "xx" (M4)
will appear. In case of free program numbers, 0 appears.
A-1
Close the lid of the centrifuge. Now proceed as described previously to set all important run parameters. If the lid
isn't closed when storing the program, in the display "RPM/RCF" (A-1), the words "FirSt" and "CLOSE Lid" (See
figure 21) flashes alternately. If you want to start the run without storing the program, in the display "RPM/RCF" (A-
1), the words "FirSt" and "PrESS StoreE" (See figure 22) flashes alternately.
M9
M4
M3
A-4
Figure. 19
A-4
M10
Figure. 20
3
4
7
3
4
2
EN-9