Type / Version; Electrical Specifications; Voltage Input - Danfoss MBT 9110 Manuel D'opération

Masquer les pouces Voir aussi pour MBT 9110:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

TYPE
Galvanic isolation
VERSION
Galvanic Isolation, Ex & ATEX

ELECTRICAL SPECIFICATIONS

Specifications range
Galvanic isolation
Supply voltage, DC
Galvanic Isolation, Ex
& ATEX
Internal consumption
Voltage drop
Isolation voltage, test / operation
Warm-up time
Common specifications
Communications interface
Signal / noise ratio
Response time (programmable)
EEprom error check
Signal dynamics, input
Signal dynamics, output
Calibration temperature
ACCURACY, THE GREATER OF GENERAL AND BASIC VALUES
Input type
Absolute accuracy
All
≤ ± 0.05% of span
BASIC VALUES
Input type
Basic accuracy
RTD
≤ ± 0.2 °C
Lin.R
≤ ± 0.1 Ω
Volt
≤ ± 10 μV
TC type: E, J, K, L, N, T, U
≤ ± 1 °C
TC type: B, R, S, W3, W5, LR
≤ ± 2 °C
EMC immunity influence
Extended EMC immunity: NAMUR NE 21, A criterion, burst
Effect of supply voltage variation
Vibration
IACS E10 Rev.7 test No.7
Max. wire size
Humidity
Dimensions
Tightness (enclosure / terminal)
Weight
ELECTRICAL SPECIFICATIONS, INPUT
RTD AND LINEAR RESISTANCE INPUT
RTD type
Min. value
Pt 100
-200 °C
Ni 100
-60 °C
Lin. R
0 Ω
4 | AQ40924028350901-000101
MBT 9110
084Z7442
084Z7443
-40 °C – 85 °C
MBT 9110, 084Z7442
7.2 – 35 VDC
MBT 9110, 084Z7443
7.2 – 30 VDC
25 mW – 0.8 W
7.2 VDC
1.5 kVAC / 50 VAC
5 min.
Loop Link
min. 60 dB
1 – 60 s
< 3.5 s
20 bit
16 bit
20 – 28 °C
GENERAL VALUES
Temperature coefficient
≤ ± 0.01% of span / °C
Temperature coefficient
≤ ± 0.01 °C / °C
≤ ± 10 mΩ / °C
≤ ± 1 μV / °C
≤ ± 0.05 °C / °C
≤ ± 0.2 °C / °C
≤ ± 0.5% of span
≤ ± 1% of span
≤ 0,005% of span / VDC
IEC 60068-2-6 Test FC
25 - 100 Hz : ± 4 g
40 - 2000 Hz : ± 10 g
1 x 1.5 mm
2
stranded wire
< 95% RH (non-cond.)
ø44 x 20.2 mm
IP68 / IP00
50 g
Max. value
Min. span
850 °C
25 °C
IEC 60751
250 °C
25 °C
DIN 43760
5000 Ω
30 Ω
© Danfoss | Climate Solutions | 2022.03
Max. offset
Cable resistance per wire (max.)
Sensor current
Effect of sensor cable resistance (3- / 4-wire)
Sensor error detection
TC INPUT
Type
B
E
J
K
L
N
R
S
T
U
W3
W5
LR
Max. offset
Cold junction compensation
Sensor error detection
Sensor error current

VOLTAGE INPUT

Measurement range
Min. span
Max. offset
Input resistance
OUTPUT
CURRENT OUTPUT
Signal range
Min. signal range
Updating time
Output signal at EEprom error
Load resistance
Load stability
SENSOR ERROR DETECTION
Programmable
NAMUR NE43 Upscale
Standard
NAMUR NE43 Downscale
Of span =
© Danfoss | Climate Solutions | 2022.03
Min. temperature
Max. temperature
400 °C
1820 °C
-100 °C
1000 °C
-100 °C
1200 °C
-180 °C
1372 °C
-100 °C
900 °C
-180 °C
1300 °C
-50 °C
1760 °C
-50 °C
1760 °C
-200 °C
400 °C
-200 °C
600 °C
0 °C
2300 °C
0 °C
2300 °C
-200 °C
800 °C
When detecting
Else
50% of selec. max. value
5 Ω
nom. 0.2 mA
< 0.002 Ω / Ω
Yes
Min. span
Standard
200 °C
IEC584
50 °C
IEC584
50 °C
IEC584
50 °C
IEC584
50 °C
DIN 43710
100 °C
IEC584
200 °C
IEC584
200 °C
IEC584
50 °C
IEC584
75 °C
DIN 43710
200 °C
ASTM E988-90
200 °C
ASTM E988-90
50 °C
GOST 3044-84
50% of selec. max. value
≤ ± 1.0 °C
Yes
nom. 33 mA
0 mA
-12 – 800 mV
5 mV
50% of selec. max. value
10 mΩ
4 – 20 mA
16 mA
440 ms
≤ 3.5 mA
≤ (V
- 7.2) / 0.023 [Ω]
supply
≤ ±0.01% of span / 100 Ω
3.5 – 23 mA
23 mA
3.5 mA
Of the presently selected range
AQ40924028350901-000101 | 5

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

84z7442084z7443

Table des Matières