Further Disassembly Options - VWR Digitrate Pro Manuel D'instructions

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 16
DIGITRATE PRO

Further Disassembly Options

To clean or replace the FEP dispense tube (12) and/or
threaded platform base (10), follow these steps;
Turn the anti-drip valve (11) to the closed position
and unscrew the threaded ring (4) to separate the
pedestal base from the glass barrel of the Digitrate Pro.
Remove dispense tube protection cover (13) as shown
oppositein an upward direction allowing FEP dispense
tube (12) to be pulled out of anti-drip valve assembly
(11) indirection indicated.
Reassemble in reverse, make sure the front end of the
FEP dispense tube (12) is clipped into the dispense tube
protection cover (13) from the underside.
Reassemble the remaining components in reverseorder,
making sure that the FEP 'O' ring (9) is well located into
its platform recess.
Note:
Bevelled inside edge of barrel must be at the top end
when fitted. The piston rack must engage in to the
electronic head with the piston rack teeth pointing
forward in the direction of the display.
After reassembling the instrument, prime with distilled
water to ensure that assembly has been correctly
followed and piston is working smoothly. Check that
no leaks occur.
10
|
VWR
INSTRUCTION MANUAL
Dispense tube
protective cover (13)
Digitate Pro wheels (1) (pair)
ASA electronic head casing (2)
Polypropylene protection sleeve (3)
Threaded platform ring (4)
|
EN
FEP dispense tube (12)
Anti-drip tap valve (11)

Publicité

Table des Matières
loading

Table des Matières