Schematic representation
Example to connect a LED on the n°1 digital output
Example to connect a relay on the n°1 digital output
(external power supply)
Note: a protective diode must be added in parallel of the relay if it isn't integrated.
The maximum voltage is 30V
and the maximum current is
60mA per output.
The supply voltage of the
optocoupler outputs are nor-
mally brought from the out-
side, the negative pole on
the
(pin 9)
The protective diode is indis-
pensable in case of inductive
load (solenoid valve, relay,
...)
R
GND's external power supply
External power supply (max. +30V)
NO
COM
8
common
transmitters
NC