To use Delta (sampling) or Mean (sampling) mode,
red, then the sampling function selected for the
desired external contact, for example pedal 1.
sed the channel value is temporarily recorded (n
indicates the number of values recorded) and the
delta or mean is calculated as a function of the
recorded values. This allows a delta to be pro-
duced at several points using a single instrument.

Channel calibration at 2 points

It is possible to calibrate the channel at two reference
points. A multiplication factor is then calculated so
allowing the measurement of diameters on V or the
calibration of a channel depending on the reference
block. This multiplication factor can then be activated
or deactivated. This calibration only affects the chan-
nel and not the instrument allocated to the channel.

System setting

NO GO channel: when the channel value is outside tolerances
GO channel: when the channel value is within tolerances
All manuals and user guides at all-guides.com
-
-
This menu enables access to the unit's
Recorded measures visualisation
Modify printing options