Calibration Mode - Tintometer Aqualytic AL100 Mode D'emploi

Masquer les pouces Voir aussi pour Aqualytic AL100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 52

Calibration Mode

GB
RESULT
CAL
Mode
Zero
Test
CAL
RESULT + x
On
Off
:
:
48
All manuals and user guides at all-guides.com
The result is shown in the display, alternating with CAL.
If the reading corresponds with the value of the calibration standard
(within the specified tolerance), exit calibration mode by pressing the
[ON/OFF] key.
Changing the displayed value:
Pressing the [MODE] key once increases the displayed value by 1 digit.
Pressing the [ZERO/TEST] key once decreases the displayed value by
1 digit.
Press the corresponding key until the reading equals the value of the
calibration standard.
By pressing the [ON/OFF] key, the new correction factor is calculated
and stored in the user calibration software.
Confirmation of calibration (3 seconds).
Note: Seperate calibration of the measuring range for Bromine is
not possible. The unit uses the calibration for the Chlorine measuring
range (CL 6).
MD100_1 02/2018

Publicité

Table des Matières
loading

Table des Matières