Tintometer Aqualytic AL100 Mode D'emploi page 42

Masquer les pouces Voir aussi pour Aqualytic AL100:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 52
Methods
GB
tA
0.0.0
Zero
Test
tA
RESULT
42
All manuals and user guides at all-guides.com
Alkalinity-m with Tablet
5 – 200 mg/l
Fill a clean vial (24 mm Ø) with 10 ml of the water sample and
perform zero calibration (see "Operation").
Add one ALKA-M-PHOTOMETER tablet straight from the foil to
the 10 ml water sample and crush the tablet using a clean stirring rod.
Close the vial tightly with the cap and swirl gently several times until
the tablet is dissolved.
Place the vial in the sample chamber making sure that the
are aligned.
Press the [ZERO/TEST] key.
The method symbol flashes for approx. 3 seconds.
The result is shown in the display in mg/l CaCO
Tolerance: ± 5% (full scale)
marks
.
3
MD100_1 02/2018