Starting your appliance

The appliance will begin start up when initially plugged in or
when power resumes after a power outage. If the appliance
has been turned off during use, "OFF" will appear on the
display. To start the appliance from the "OFF" position press
and hold the ON/OFF button for three seconds.

Set temperature

To set the set temperature, press and continue to hold
"SET" button. After one second, set temperature will be
displayed. While holding "SET" button use the "WARMER"
or "COLDER" button to desired set temperature.
NOTE
Momentarily pressing & releasing "SET" button will access
information menu of control. Refer to appendix A for infor-
mation on this feature.

Appliance operation

The available temperature range of the refrigerator is 2° to
8° C.
It may take up to 24 hours for your appliance to reach de-
sired temperature. This will depend on amount of contents
loaded and number of openings and closings of the door.
For best results allow the appliance to "pull down" to de-
sired set temperature before loading. Once contents are
loaded, allow at least 48 hours for temperature to stabilize
before making any adjustments to the set temperature.

Alarms

Your MicroSentry II™ appliance control will monitor ap-
pliance function and alert you with a series of audible and
visual alarms. For all alarm and error codes, the alarm
LED located in the top left corner of the display panel will
illuminate a steady amber. For all alarm and error codes
except condenser cleaning, an audible alarm will sound,
beeping in one (1) second intervals.
Door Ajar Alarm: If the door has been
left open for over five (5) minutes, the alarm
will sound and the display panel will flash
"do." The alarm will stop once the door has been closed.
• High and Low Temperature Alarm: If
your unit reaches an unacceptable tempera-
ture, above or below your set temperature for
a period of greater than 60 minutes, the tem-
perature alarm will sound in one (1) second
intervals. The display panel will flash either "hi" or "Lo" de-
pending upon the condition. "hi" indicates a high tempera-
ture alarm, and "Lo" indicates a low temperature alarm.
USING THE MICROSENTRY II ™ CONTROL
Alarm
Set
Colder
Warmer
Press and Hold
The factory preset for the "hi" and "Lo" temperature devia-
tion is +2°C from set temperature for the "hi" and -2°C from
set temperature for the "Lo". If these temperature variations
need to be adjusted, please call AGA Marvel customer
service at 800-223-3900 for instructions.
• Temperature Sensor Fault: If the control
detects that the temperature sensor is not
properly functioning, an audible alarm will
sound, and "E1" will flash on the display. Please call AGA
MARVEL customer service or your dealer if this error code
is displayed.

Alarm Mute

Press any key to mute the audible portion of an alarm.
NOTE-This action will only mute the alarm. If the condi-
tion that caused the alarm continues, the alarm code will
continue to flash and will sound for 20 seconds every 60
minutes.

Turning Appliance Off

To turn appliance off, press and hold "ON/OFF" button for
three (3) seconds. "OFF"
display.
Additional MicroSentry II™ Features
Refer to Appendix A for details on additional features avail-
able.
7
F C
Figure 8
Control located
on face of door
ON/OFF
Press and Hold
will appear on the

Publicité

Chapitres

Table des Matières
loading

Ce manuel est également adapté pour:

Marvel scientific ms24ra serieMarvel scientific sa24ra serieMarvel scientific marvel scientific ms15raMarvel scientific ms15rasMarvel scientific ms24raMarvel scientific ms24rag ... Afficher tout

Table des Matières