SEFRAM 3515 Manuel D'utilisation page 57

Masquer les pouces Voir aussi pour 3515:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

All manuals and user guides at all-guides.com
(2) Current
3515
Function
Range
99.99A
50 ~ 60Hz ± (1.5% + 5dgt) **
ACA
>60 ~ 400Hz ± (2% + 5dgt) **
599.9A
0.10A ~ 99.99A
LPF
50 ~ 60Hz ± (1.5% + 5dgt) **
ACA
>60 ~ 400Hz ± (5% + 5dgt) **
599.9A
** The measured value <1000dgt, add 5 dgt to the accuracy.
3525,3535
Function
Range
99.99A
DCA
599.9A/999.9A*
0.10A ~ 99.99A
50 ~ 60Hz ± (1.5% + 5dgt)
ACA
>60 ~ 400Hz ± (2% + 5dgt)
599.9A/999.9A*
0.10A ~ 99.99A
LPF
50 ~ 60Hz ± (1.5% + 5dgt) **
ACA
>60 ~ 400Hz ± (5% + 5dgt) **
599.9A/999.9A*
* 3525 : 599.9A ; 3535 : 999.9A
** The measured value <1000dgt, add 5 dgt to the accuracy.
Overload protection : 1000A
for 3535
rms
600A
for 3515,3525
rms
Position Error : ±1% of reading.
AC Conversion Type and additional accuracy is same as AC
Voltage.
AC+DC Arms Accuracy : Same as ACA spec + DCA spec.
- For better measurement accuracy of high current, do not
measurement more than 10min.(for 3515)
- DCA affected by the temperature and the residual magnetism.
Press HOLD key > 2sec to compensate it.
26
Accuracy
Accuracy
± (1.5% + 0.2A)
± (1.5% + 5dgt) **
**
**

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

35253535

Table des Matières