PRIOR TO USE
• Fully charge the battery before using scale for the
first time.
• Let the scale warm up for 15 minutes before use.
WARNING! Never drop items onto the platform.
The scale is a sensitive precision instrument.
WARNING! Do not exceed the scale capacity
and overload the scale.
WARNING! Do not stack items on the scale
platform when the scale is not in use.
TURNING THE SCALE ON/OFF
ON/OFF key is located in the lower right hand corner of
the display.
CAUTION! Never power the scale on with weight
on the platform.
1. To turn the scale on, press the ON/OFF key
The scale will show the version and enter a self-test
mode. After completing the self-test, the scale will
enter the weighing application mode.
2. To turn the scale off, press the ON/OFF key
CHANGE UNIT OF MEASURE
Press the UNIT key
appears – lb, oz or kg.
CALIBRATION
The scale is pre-calibrated during production. You
must use a precision test weight to properly calibrate
the scale. The test weight CANNOT exceed the scale's
capacity.
NOTE: Calibration can be done in English (lb.)
and Metric system weight (kg) at the scale's full
capacity.
CALIBRATION WEIGHTS
Model #
H-5836
H-5837
PAGE 5 OF 36
until the desired weighing unit
Lb.
100 lb.
50 kg
200 lb.
100 kg
OPERATION
SINGLE POINT CALIBRATION (SPAN)
Single point calibration uses two calibration points, zero
and full weight capacity of the scale.
1. Turn on the scale.
2. Press the PCS/SET key
3. Display will be show
4. Press the G/N key
and then the TARE key
5. Press the TARE key
6. Press the ZERO key
TARE key
7.
Press the ZERO key
TARE key
.
8. Press the ZERO key
9. Press the ZERO key
.
will show
10. Remove all the weight from the platform.
11. When indicator is stable, press the ZERO key
to confirm.
12. Display will show the last calibration weight. Change
the calibration weight value to the full capacity of
the scale. Press the G/N key
ACCT/TTL/Print key
and press the TARE key
13. When the calibration value is correct, press the ZERO
key
14. Display will show
15. Place the calibration weight on the platform.
Kg
16. When indicator is stable, press the ZERO key
to confirm.
17. Display will return to weighing mode.
during the self-test.
pn.
, then the UNIT/ESC key
. Display will show
until display shows
to confirm. Press the
until display shows
to confirm. Press the
cal.
to select
to confirm.
to enter calibration, display
unld
.
to change the active digits
to change the value.
to confirm.
load
.
.
p 2 mod
.
sigr.
and the
key
0621 IH-5836