Description of function
With the InTrac
®
787 retractable housings, electrodes can be
cleaned or replaced without interrupting the ongoing process.
The process can be closed safely with the integrated ball valve.
The integrated flushing port allow a rinsing of the electrode in
the housing.
The immersion tube is moved into the desired end position
("Measuring" or "Maintenance") by hand.
– In the "Measuring" position, the immersion tube is fully in-
serted in the process.
– In the "Maintenance" position, the electrode can be cleaned
or removed/exchanged. The ball valve seals the housing
against ingress of process medium.