InTrac
®
787
4.2.2
Fitting the electrode/sensor
1
3
5
1. Pull out the immersion tube.
2. Close the ball valve.
3. Disconnect the immersion tube right at the ball valve.
4. Disconnect the electrode holder by turning it counterclock-
wise.
5. Disconnect the electrode cable.
6. Remove the electrode.
7. Check washers and O-rings on electrodes for damage
and replace if necessary.
8. Remove watering cap from electrode tip and rinse electrode/
sensor with water.
9. Carefully insert the electrode into the electrode holder and
screw in by hand as far as it will go. Important: use no tools!
10. Re-Install the complete housing.