1. Instructions

TRI-SCALER Compact is used for the scaling of teeth. It is also an indispensable aid for the prevention and treatment of perio-
dontal disease. The new product TRI-SCALER Compact has scaling, periodontal, and endodontic functions.
It contains the following features:
- Automatic frequency tracking ensure that the machine always works on the best frequency and performs more steadily.
- The handpiece is detachable and can be autoclaved to a high temperature of 135°C and a pressure of 0.22MPa.

2. Components

1. Main unit
2. Detachable handpiece
3. Power supply
4. Foot pedal
5. Scaling tip
6. Water line
7. Torque wrench
8. Endodontic wrench
9. Endochuck
10. File
11. Instruction manual

3. The main technical specifications

• Transformer input : 220 V - 240 V~ 50 Hz/60 Hz 150mA
• Main unit input: 24 V~ 1,3 A
• Output primary tip vibration excursion : ≤ 100 μm
• Output tip vibration frequency : 28 kHz ±3 kHz
• Output half-excursion force : <2 N
• Output power : 3 W to 20 W
• Main unit fuse : T 1,6 AL 250 V
• Power source fuse : T 0,5 AL 250 V
• Water pressure: 0.1 bar to 5 bar (0.01 MPa to 0.5 MPa)
• Weight of main unit : 0.65kg
• Weight of transformer : 1 kg
• Operation mode : continuous operation
• Type of protection against electric shock : Class II equipment
• Degree of protection against electric shock : Type BF applied part
• Degree of protection against harmful ingress of water : Ordinary equipment (IPX0)
• Degree of protection against water (used on the foot switch): IPX1
• Degree of safety of application in the presence of a Flammable Anesthetic Mixture with air or with Oxygen or Nitrous Oxide :
Equipment not suitable for being used in the presence of a flammable anesthetic mixture with air or with oxygen or nitrous
oxide.