The interval for the data transmission (Int 2) determines the chronological
interval between automatic data transmissions. After the selected interval
expires, the current data record is transmitted to the interface. To set up the
transmission interval, press <RCL> while holding down the <RUN/ENTER>
key (display Int 2). Then, set the interval with <L> <M>.
Recorder (Analog output)
Via the analog output, you can transmit data to a recorder. Connect the
analog output to the recorder using the AK323/S interface cable.
The data output automatically switches to Recorder output.
Note
The output on the analog output corresponds to the value shown on the
display (pH 7.00 is output as 700 mV).
Configuring
Note
You can leave the configuration menu at any time with <M>. The parameters
that have already been changed are stored.
Procedure for configuring (factory settings appear in bold typeface):
l Switch off the measuring instrument.
l Press <ON/OFF>> while pressing the <M> key.
l Select the required Baud rate with <L> <M>.
Selection: 1200, 2400, 4800, 9600 Baud.
l Then press the <RUN/ENTER> key.
l Select the required display during the pH calibration with <L> <M>.
Selection: Buffer nominal value (/pH) or current electrode voltage (mV).
l Then press the <RUN/ENTER> key.
l Set the required calibration interval with <L> <M>.
Selection: 1 ... 7 ... 999 d.
l Then press the <RUN/ENTER> key.
l Select the date and time step-by-step with <L> <M>. In between, press
the <RUN/ENTER> key each time.
l When the last <RUN/ENTER> key has been pressed, the instrument
switches to the last active measuring mode.
pH 197i