l Immerse the pH electrode in the second buffer solution to continue the
two-point calibration.
l When measuring without a temperature sensor:
Enter the temperature of the third buffer solution with <L> <M>.
l Press <RUN/ENTER>. The AutoRead measurement begins. The
electrode voltage (mV) or the buffer nominal value appears on the
display.
l When the measured value is stable, the instrument shows the value of
the slope and the evaluation of the three-point calibration.
l Press <RUN/ENTER>. The instrument displays the value of the
asymmetry.
l Switch to the measuring mode with <M>.
Calibration