Press the
button for 1 second to switch the thermometer on.
Following a successful self test, the device emits two short beeps.
•
Remove the cap by gently pushing it up (1) and then pulling it off
forwards (2).
24
All manuals and user guides at all-guides.com
Make sure that the sensor tip and also the ear canal are clean.
As the ear canal is slightly curved, you have to pull the ear slightly up and
backwards before inserting the sensor tip. This is important so that the sensor
tip can be pointed directly at the eardrum.
•
Insert the sensor tip carefully and press the
•
Release the
button. The end of the measuring time is signalled with a
short beep and the measured value appears on the display.
If the measured value is within the normal range (< 100.4°F/38°C), the green
LED lights up for 3 seconds. If the measured value is higher (≥ 100.4°F/38°C,
high temperature), the red LED is illuminated.
2
1
button for 1 second.