2.3 Technical data

2.3.1 Product-specific data
Jar sizes:
2.3.2 Dimensions and weight
TOPITEC
AUTOMATIC (device dimensions) including touch screen
®
Width:
Depth:
Height:
Total weight:
2.3.3 Power supply / electrical parameters
Operating voltage:
Phases:
Protective earth conductor: PE (yellow/green)
Frequency:
Protection rating:
Overvoltage category:
EMC:
Protection class:
Installation standard:

Power supply

Fuses:
Rated power:
12
TOPITEC
®
AUTOMATIC User Manual | 12-2020
Jars up to max. diameter 118 mm
Single-patient containers, single-patient jars, jars for stock preparation
approx. 240 mm
approx. 240 mm
approx. 680 mm
approx. 18 kg
1 x 230 V
AC
±10%
1 Ph / PE
50 Hz ± 1%
I
2
Laboratories, domestic environments, SMEs
IP 20
DIN EN 61010-1: 2011
1x 10A slow-acting and 1x 1A slow-acting
approx. 400 W
Warning
The device must be connected to a safety socket. The mixing system is equipped
with a fine protection fuse. Medium and general protection must be provided as
part of the power distribution system within the building.
The device is protected by 2 fuses (1x 10A slow-acting and 1x 1A slow-acting).
The fuse holder is located on the back. At the location of operation, it must be
protected by a 10A circuit breaker (characteristic B). When replacing a fuse, use
one of the same type, rating and characteristic.