During cable installation, please note the following:
• Use only shielded and low−capacitance measurement cables (HBM cables
meet these conditions).
• Do not lay measurement cable parallel to high-voltage power lines or
control circuits. If this is not possible, protect the measurement cable with
armoured steel tubing, for example. Keep it at least 50 cm away from the
other cables.
• Avoid stray fields of transformers, motors and contactors.
• Do not earth transducer, amplifier and display device more than once. All
the devices in the measuring chain are to be connected to the same
earthed conductor.
Six-wire connection
Four-wire connection
WH
BK
RD
BL
YE
Feedback bridges in the transducer
Fig. 5.2: Connection in six−wire and four−wire techniques
HBM
WH (white)
Measurement signal (+) U
BK (black)
Excitation voltage (−) U
RD (red)
Measurement signal (−) U
BL (blue)
Excitation voltage (+) U
GN (green)
Sensor circuit (+)
GY (grey)
Sensor circuit (−)
Shield
Cable shielding, connected to housing
WH (white)
BK (black)
RD (red)
When the transducer is connected
to an HBM measuring amplifier in
BL (blue)
accordance with the specifications
GN (green)
in the table, a compressive load
will be displayed with a positive
GY (grey)
sign. To change the polarity, swap
YE (yellow)
the red and white wires.
U1A
A
B
A
B
a0279-4.0 de/en/fr