Warranty:
The breast pump is guaranteed for 3 years from the date of purchase against any manufacturing defect,
except case of misuse.
For after sales service under warranty, please contact DTF MEDICAL or your local distributor.
The technical documentation is available on request.
Recycling:
The device must be disposed of in accordance with local regulation on the disposal of electrical
appliances. So you contribute to the human health and environmental protection. Product on the
market since August the 13
Device class (breast pump)
Device class in electrical safety (breast pump)
Type of applied parts
Protection level
Expected shelf life of accessories
Expected shelf life of autoclavable accessories
Expected shelf life of the breast pump
Collection container capacity
Weight
Dimensions
Voltage
Consumption
Fuses
Suction values
Cycles
Sound level
Running mode
Power supply
Ambient temperature (T) - For transport and storage
Ambient temperature (T) - When in functioning
Relative humidity (H)
For transport, storage and functioning
Atmospheric pressure (P)
For transport, storage and functioning
Essential performance features:
•
Provide adequate suction so that mothers can express their milk.
•
Control the rate of suction to best meet the baby's sucking rhythm.
Applied part: Breast shield.
Identification:
Each breast pump has a unit identification (reference and number) which indicates, among other things, the
exact date of manufacture and the serial number of the breast pump. The first two digits show the year of
manufacture (e.g. 68: 1950 + 68 = 2018) and the next three digits indicate the day of the year of manufacture
th
(e.g. 285 = 12
October). The last two digits indicate the serial order number in the series. In addition, each
accessory is identified by a lot number mentioned. For more information, refer to the instructions for use of
the KIT EXPRESSION.
9.
Warranty / Recycling
th
2005.
10. Technical specifications
IIa
II
BF
IP 30
1 year
50 cycles
5 years
100ml
4 kg
L360 x l250 x h180mm
Available in 230V – 50Hz; 220V – 50Hz or 220V – 60Hz
100VA
T 0,5A H 250V
Variable from -50 to -300hPa
40 to 125 cycles/minutes
<57dBa
Continuous
Alternative current 230V – 50Hz
22
-25°C < T < +70°C
+5°C < T < +40°C
15% < H < 93%
700 hPa < P < 1060 hPa