Declaration – electromagnetic emissions and immunity
for equipment and systems that are not life-supporting and are specified for use only in a shielded location
Declaration – electromagnetic immunity
This Wi-Fi Smart Blood Pressure Monitor is intended for use in the electromagnetic environment specified below. The customer or the user of the
Wi-Fi Smart Blood Pressure Monitor should ensure that it is used in such an environment.
Immunity test
Electrostatic discharge
(ESD)IEC 61000-4-2
Electrical fast transient/ burst
IEC 61000-4-4
Surge IEC 61000-4-5
IEC 60601 test level
Contact: ±8 kV
Air: ±2 kV, ±4 kV, ±8 kV,
±15 kV
2 kV for power supply lines
1 kV for input/output lines
1 kV differential mode
2 kV common mode
Compliance level
Electromagnetic environment – guidance
Contact: ±8 kV
Floors should be wood, concrete or ceramic tile. If floors are
Air: ±2 kV, ±4 kV, ±8 kV,
covered with synthetic material, the relative humidity should
±15 kV
be at least 30%.
N/A
The main power quality should be similar to that of a typical
commercial or hospital environment.
ine(s) to line(s): ±1 kV
The main power quality should be similar to that of a typical
100 kHz repetition
commercial or hospital environment.
frequency