Accuracy Table
(7) AUTO V (AC/DC voltage automatic detection)
During AC detection: Conforms to accuracy specifications described in "(10) AC+DC voltage" (p. 51).
During DC detection: Conforms to accuracy specifications described in "(9) DC voltage" (p. 49).
(8) AC voltage
Measurement value/MAX/MIN/AVE
Range
Accuracy
(Auto-range
guarantee range
threshold)
(Resolution)
6.000 V (more
0.000 V to 0.299
than 6000
V (0.001 V)
counts)
0.300 V to 6.000
V (0.001 V)
46
Accuracy guarantee
1
2
frequency range *
*
15 Hz ≤ f < 45 Hz
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
15 Hz ≤ f < 45 Hz
45 Hz ≤ f ≤ 66 Hz
66 Hz < f ≤ 1 kHz
Measurement accuracy
Filter OFF
Filter ON
±1.5% rdg.
±2.0% rdg.
±0.015 V
±0.015 V
±0.9% rdg.
±1.4% rdg.
±0.013 V
±0.013 V
±1.5% rdg.
±0.015 V
±1.5% rdg.
±2.0% rdg.
±0.005 V
±0.005 V
±0.9% rdg.
±1.4% rdg.
±0.003 V
±0.003 V
±1.5% rdg.
±0.005 V
Input
3
impedance *
Ω
3.2 M
±5%
Ω
3.2 M
±5%