0rAnG
MANUAL ZERO RANGE
By pressing ZERO one sets the weight percentage for the manually acquiring the gross zero; if, in the normal
weighing phase, by pressing ZERO, a weight within the programmed percentage of the total capacity, it will be
cleared.
Press TARE to select one of the suggested percentages: +/- 2, 4, 10, 20, 50, or 100 % of the max. capacity.
Confirm with ZERO or press PRINT/M+ to exit without confirming.
SPEED
SETTING ADC CONVERTER READING SPEED
By pressing ZERO one sets the reading speed of the ADC converter; the set value corresponds to the number
of readings made in a second.
Press TARE to select one of the suggested reading speeds: 7.5, 15, 30, 60.
Confirm with ZERO or press PRINT/M+ to exit without confirming.
7.8.2 "P2 CAL" – CONFIGURATION AND CALIBRATION (*)
Premise: through this step it's possible to configure and calibrate a single cell scale, or a scale having various
cells equalised externally through the junction box; the scale is connected on the only channel on the indicator.
The unit of measure of the calibration is fixed at kilogram ("kg").
(*) In case of approved instrument the parameter is read only.
DECi
NUMBER OF SCALE DECIMALS
By pressing ZERO one selects the number of scale decimals.
Press TARE to select one of the suggested values: 0.0 (a decimal),0.00 (two decimals), 0.000 (three
decimals), 0.0000 (four decimals), 0 (no decimal); confirm with ZERO.
Confirm with ZERO or press PRINT/M+ to exit without confirming.
inC
MINIMUM SCALE DIVISION
By pressing ZERO one selects the minimum scale division.
Press TARE to select one of the suggested values: 1, 2, 5, 10, 20, and 50.
Confirm with ZERO or press PRINT/M+ to exit without confirming.
CAP
SCALE CAPACITY
By pressing ZERO one selects the total capacity of the scale.
-
Use the F key to clear the value, the TARE key to increase the digit value, the U key to go on the previous
digit and the SMPL key to go on the following digit.
-
Confirm the entered value with ZERO or press PRINT/M+ to exit without confirming.
CAL
SCALE CALIBRATION
-
Press ZERO, the display shows "UNLOAD";
-
Unload the scale and press ZERO when the stability led is on;
-
Set the sample weight value by using the F key to clear the value, the TARE key to increase the value of
the digit, the U key to go on the previous digit and the SMPL key to go on the following digit. Confirm the
entered value with ZERO, the display shows "LOAD";
-
Put the sample weight on the scale and press ZERO when the stability indicator is on;
-
If the calibration is correct, the display shows "PASS" for an instant, and the indicator returns to weighing;
-
If the calibration is wrong, the display shows "FAIL" for an instant and then "CAL".
15