ProMed EMT-4 Instructions D'utilisation page 60

Publicité

Les langues disponibles
  • FR

Les langues disponibles

manual_emt-4_270607
27.06.2007
60
D
G
ISPLAY
UIDE
Program display
Frequency display
Pulse width
display
Intensity,
channel 1
When the unit is switched on, all readouts are active for a short ti-
me before the unit switches to standby (wait) mode.
14:44 Uhr
Seite 60
Pause
Function check
Treatment time
Intensity,
channel 2
P
S
...
ROGRAM
ELECTION
Preset programs P-01 to P-11 and memory function
Press the "ON/OFF" button to switch on the unit.
1.
Keep pressing the "Program" button until one of the preset pro-
2.
grams (P-01 to P-11) is shown on the display.
Press the "Set" button. The time will flash on the display, " ´ "
3.
for minutes and " `` " for seconds. Use the "+" or "–" buttons
to enter the desired treatment duration (20 minutes to 2 hours)
in steps of ten minutes.
Press the "Select" button 2 times. The intensity display "CH1"
4.
will start to flash. Use the "+" or "–" buttons to enter the desi-
red value (from 1 to 50). Press the "Select" button once more.
The intensity display "CH2" will start to flash. Once again, use
the "+" or "–" buttons to set the desired value.
To begin treatment, press the "Select" button once more. The
5.
clock symbol at the top right of the display will start to flash.