Electromagnetic Compatibility Information
Table 1
For all ME EQuIPMEnT and ME sYsTEMs
Guidance and manufacture's declaration – electromagnetic emissions
The BC 16 is intended for use in the electromagnetic environment specified below.
The customer or the user of the BC 16 should assure that it is used in such an environment.
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
Harmonic emissions IEC 61000­3­2
Voltage fluctuations/flicker emissions
IEC 61000­3­3
Table 2
Guidance and manufacturer's declaration – electromagnetic immunity
The BC 16 is intended for use in the electromagnetic environment specified below. The customer or the user of the BC 16 should assure that it
is used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 61000­4­2
Electrical fast transient/burst
IEC 61000­4­4
Power frequency (50/60 Hz)
magnetic field IEC 61000­4­8
noTE: U
is the a.c. mains voltage prior to application of the test level.
T
Compliance
Electromagnetic environment – guidance
Group 1
The BC 16 uses RF energy only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in nearby electronic equipment.
Class B
The BC 16 is suitable for use in all establishments, including domestic establish­
ments and those directly connected to the public low­voltage power supply network
Not applicable
that supplies buildings used for domestic purposes.
Not applicable
IEC 60601 test level Compliance level Electromagnetic environment – guidance
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for power
± 2 kV for power
supply lines
supply lines
± 1 kV for input/
N/A
output lines
3 A/m
3 A/m
Floors should be wood, concrete or ceramic tile. If floors are covered
with synthetic material, the relative humidity should be at least 30 %.
Mains power quality should be that of a typical commercial or hos­
pital environment.
Power frequency magnetic fields should be at levels characteristic
of a typical location in a typical commercial or hospital environment.
86