8. Safety features
EDDYTHERM 2x is equipped with the following safety features:
Automatic overheating protection

Automatic current control

In the TEMP MODE the heater will switch off if the temperature probe

does not register a temperature increase of 1° every 15 seconds. To
increase the interval to 30 seconds, press MODE and DOWN at the
same time.
EDDYTHERM 2x is equipped with a main switch with over-current

circuit breaker.
9. Troubleshooting
A system fault will be indicated by an acoustic signal and one of the
following fault codes on the main display:
Display
Fault
E01 E
General system failure
E02 E
Memory failure
E03 E
Overheated coil
E05 E
Temperature increase of less than 1°C
every 15 seconds (or 1°C every 30
seconds)
E06 E
Temperature probe not connected (or
defective)
E07 E
Failure during current measurement
E08 E
Failure during communication with
printed circuit board
E09 E
Overheated printed circuit board
10. Maintenance
EDDYTHERM 2x may only be opened by authorized PRÜFTECHNIK service
partners.
16
Action
Return EDDYTHERM 2x for repair
Return EDDYTHERM 2x for repair
Wait until the inductive coil cools
Check the temperature probe connection. If the
connection is OK, select the 30 second interval as
described in section 6 or operate the heater in TIME
MODE
Check the temperature probe
Return EDDYTHERM 2x for repair
Return EDDYTHERM 2x for repair
Wait until the printed circuit board cools.