Putting Into Operation; Select Measuring Range; Bridge Zero Balancing; Mr Fine Adjustment - HBM ME50S6 Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12
3

Putting into Operation

If the manufacturer's settings, such as the coarse measuring range, are re-
tained the operation is limited to the following points:

3.1 Select Measuring Range

With the factory settings measuring range 1 (" 1 mV/V) is always switched
on. Measuring range 2 (calibration at "0.2 mV/V) can be selected either in-
ternally via DIP switch 24/4 or externally by connecting a single-pole change-
over switch to pin 25c (see chapter 4.2).

3.2 Bridge Zero Balancing

The bridge zero balancing is carried out by means of the screwdriver poten-
tiometers P23 (fine) and switches S22 (coarse) and S23 (polarity). (See also
chapter 4.4).

3.3 MR Fine Adjustment

The fine adjustment of the measuring range is made seperately for both
ranges by means of the screwdriver potentiometers P21 (measuring range 1)
and P22 (measuring range 2). The potentiometers, which are accessible from
the front, permit corrections of approx. 35 % (related to the set measuring
range). As far as measuring range 1 is concerned, the fine adjustment is also
possible via an external potentiometer (for this see chapter 4.3).
NOTE
P22 must not be altered in order not to detune the manufacturer's setting
(MR2 " 0.2mV/V) (see chapter 4.5).
HBM
ME50S6
A0260−3.0 en/de/fr

Publicité

Table des Matières
loading

Table des Matières