Annex
Guidance and manufacturer´s declaration – electromagnetic immunity
The device or system is intended for use in the electromagnetic environments listed below. The customer and/or user
of the device or system must ensure that it is used in an electromagnetic environment as described below:
Immunity
IEC 60601
test
– test level
EN
Conducted
3 Vrms
RF IEC
150 kHz to
61000-4-6
80 MHz
Radiated
3 V/m
RF IEC
80 MHz to
61000-4-3
2.5 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2:
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorpti-
on and reflection from structures, objects and people.
Compliance
Electromagnetic environment - guidance
level
Portable and mobile RF communications equipment should be used no
closer to any part of the device, including cables, than the recommended
separation distance calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance:
3 Vrms
d= 1.2 √P
d = 1.2 √P
3 V/m
80 MHz to 800 MHz
d = 2.3 √P
800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts
(W) according to the transmitter manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electro-
magnetic site survey,
frequency range.
Interference may occur in the vicinity ofequipment marked
with the following symbol:
64
should be less than the compliance level in each
a
b