Annex
Guidance and manufacturer´s declaration – electromagnetic emissions
The device is intended for use in the electromagnetic environments listed below, and should only beused in such
environments:
Emissions test
RF emissions CISPR 11
RF emissions CISPR 11
EN
Harmonic emissions IEC 61000-3-2
Voltage fluctuations / flicker emissions IEC
61000-3-3
Guidance and manufacturer´s declaration – electromagnetic emissions
The device or system is intended for use in the electromagnetic environments listed below. The customer and/or user
of the device or system must ensure that it is used in an electromagnetic environment as described below:
Immunity test
Electrostatic discharge
(ESD) IEC 61000-4-2
Power frequency (50/60 Hz)
magnetic field
IEC
61000-4-8
Compliance Electromagnetic environment - guidance
Group 1
Class B
Class A
complies
IEC 60601 –
Compliance
test level
level
± 6 kV
± 6 kV contact
contact
± 8 kV air
± 8 kV air
3 A/m
3 A/m
62
RF energy is used only to maintain device's operation.
Therefore, its RF emissions are so low that it's not
likely to cause any interference in nearby electronic
equipment.
The device is suitable for use in all establishments,
including domesticestablishments, and those directly
connected to the public low-voltage power supply net-
workthat supplies buildings used for domestic purposes.
Electromagnetic environment - guidance
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30 %.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typicalcommer-
cial or hospital environment.