5. Setting up

5.1 Delivery

The nursing home bed is generally completely assembled when delivered, or is assembled by
specialists on site.
Check the delivery for completeness by consulting the delivery papers.
Write down potential defects or damage on the delivery note.
Inform your responsible service partner immediately about possible transport damage or
defects. The address and phone number are noted on the last page of this manual.

5.2 Setting up the bed

Make sure that the floor surface is suitable, so that the floor is not damaged. Not suitable
are floors that are, for example, too soft, unsealed or with imperfections. Soft wooden floors,
porous and soft stone floors, carpets with foam backing, soft linoleum or similar flooring are
mostly unsuitable. If in doubt, please contact your flooring company and the technical custom-
er service of wissner-bosserhoff GmbH.
Ensure that the bed stands freely in the room so that there is a sufficient safety clearance all
around the bed for all adjustment positions (including the tilting function). For the extreme
positions of the bed, an additional safety clearance of 2.5 cm from objects, walls and installa-
tions must be taken into account, as protection from entrapment. The safety clearance of the
head end from objects, walls and installations is at least 20 cm.
Check for the existence of mechanical and electrotechnical protective measures, such as
wall-mounted bumpers and corner guards, RCB switches etc.
Pay attention to the whereabouts of useful wall sockets for the power connection for the bed.
Unsuitable are positions that could lead to collisions or stretching when the bed is adjusted, or
in which the power cable stretches too far across the floor or even under the bed.
In general, the use of extension cables or multiple adapters that may lie unsecured on the
floor of the room should be avoided . For the power source, a 220/230 volt, 50 Hz power
socket in accordance with VDE regulation must be used (VDE = German Electrical Engineering
Association).
Magnetic objects or magnetic fields must be kept at a distance.
The nursing home bed has to be positioned in the room in a way that it can be easily
disconnected from the power supply.
The nursing home bed is usually not equipped with an equipotential connection. A
combination with other electro-medical equipment is therefore not permitted with-
out further safety measures. An equipotential connection can be optionally provid-
ed, if required. For further information, get in touch with the technical customer
service.