Pari Montesol 177 Notice D'utilisation page 43

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 73
Using a standard thermal disinfector for baby
bottles (not a microwave oven)
EQUIPMENT:
– Thermal disinfector with a runtime of at least 6 minutes.
PROCEDURE:
CAUTION
Risk of infection due to inadequate disinfection
Inadequate disinfection encourages the growth of bacteria
and thus increases the risk of infection.
• Make sure that the disinfector is clean and operating prop-
erly before every disinfection process.
• Allow the disinfection to continue until the disinfector
switches off automatically or the minimum disinfection
time stated in the instructions for use of the disinfector has
elapsed. Do not switch the device off prematurely.
CAUTION
Risk of infection due to moisture
Moisture encourages the growth of bacteria.
• Remove all parts from the disinfector as soon as the disin-
fection process is finished and allow them to dry.
Regarding the performance of the disinfection, the duration of
the disinfection procedure and the quantity of water required,
follow the instructions for use of the disinfector you are using.
DRY:
• After the disinfection process is complete, place all parts on
a dry, clean and absorbent surface and allow them to dry
completely.
– 19 –

Publicité

Table des Matières
loading

Table des Matières