KaVo DEXIS Titanium Manuel De L'utilisateur page 46

Capteur
Masquer les pouces Voir aussi pour DEXIS Titanium:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

portable and mobile RF communications equipment and the DEXIS Titanium sensor
The DEXIS Titanium sensor is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The cus-
tomer or the user of the DEXIS Titanium sensor can help prevent electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the DEXIS Titanium sensor as recommended below, according to the
maximum output power of the communications equipment.
Rated maximum output
power
of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres (m) can be estimated
using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W)
according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-
tures, objects and people.
40
Recommended separation distances between
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
d
3
/ 5 .
V
) 1
P
0.12
0.37
1. 17
3. 69
11.67
m
d
3
/ 5 .
E
) 1
P
0.12
0. 37
1. 17
3. 69
11.67
800 MHz to 2,5 GHz
d
/ 7
E
) 1
P
0.23
0. 74
2. 33
7.38
23.33
English

Publicité

Table des Matières
loading

Produits Connexes pour KaVo DEXIS Titanium

Table des Matières