Dental phantom images should be acquired with fixed settings for X-ray
dose (kV, mA, distance), and a fixed and reproducible alignment of the
X-ray source and the dental phantom to the DEXIS Titanium sensor.
Please note:
Contact DEXIS technical support for recommendations about
commercially available dental phantoms.
Display Image
Refer to the software manual for guidance on how to ensure good display
settings and image display properties. It is recommended to view images
on an LCD or LED monitor with at least 1,920 x 1,080 pixel resolution at 0.25
to 0.26 dot pitch or better and 24 bit true color (32 bit preferred), in native
resolution.
20