Proper Care Of The Dexis Titanium Sensor Holders - KaVo DEXIS Titanium Manuel De L'utilisateur

Capteur
Masquer les pouces Voir aussi pour DEXIS Titanium:
Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CAUTION
Do NOT use any of the cleaning agents listed above for previous
versions of DEXIS sensors except CaviCide.
The slightly different design of the cable attachment and color on
those sensors will help you to distinguish them from the DEXIS
Titanium sensors. In addition, with the DEXIS Classic, connection to
the PC is through a controller. (The DEXIS Titanium and Platinum
sensors connect directly to one of the USB ports on the PC.)
Please note: DEXcare and Basic Care service level agreements do not
cover visible damage to any of the DEXIS hardware components. Abuse
or misuse of DEXIS components could result in voiding the warranty
and DEXcare or Basic Care coverage.
2.8

Proper Care of the DEXIS Titanium Sensor Holders

Your sensor holders should be disinfected after each use. The DEXIS Tita-
nium sensor holders can be disinfected in a steam autoclave using dis-
tilled water at temperature of 275
However, certain restrictions apply:
• Run the appropriate autoclave cycle according to ANSI/AAMI ST79 &
A1 & A2 & A3: 275
o
F (135
minutes.
• Always avoid direct contact of the holders with metal trays, instruments,
and heating elements.
• Always place holders in autoclave bags.
• Always situate bagged holders into the autoclave so that they are fur-
thest from the heat source.
• Do not use a chemical autoclave.
Exposing the holders to hot metal and placing them close to heating ele-
ments will reduce the lifetime of the holders.
English
F (135
C) at 216 kPa for 10 minutes.
o
o
o
C) at 216 kPa for 10 minutes, drying time 30
Install ation, Care, an d Mai nte nance
17

Publicité

Table des Matières
loading

Produits Connexes pour KaVo DEXIS Titanium

Table des Matières