6.5.1
Start/Stop
The Start/Stop measuring method starts the recording of the measured
values after tapping on the Start/Stop button.
Recording is stopped by tapping on the Start/Stop button again.
The Start/Stop measuring method implements an automatic data
compression feature and requires no further adjustment.
In addition to the current measured value, the Min and Max values are
also saved.
The measurement is completed independent of the current storage rate
at a scanning rate of 1 ms.
If the predefined memory area is full, the data compression is activated.
The storage rate is doubled and every second measured value is deleted.
In this case, the Min and Max values are also compared and the highest
and lowest measured value are transferred to the Min and Max value
respectively.
INFORMATION
The Min and Max values are not lost but are saved, even in the case
of long-term measurements.
Operation
67