3. SPECIFICATIONS
Parameter
Model Number
Power Requirements
Lamp Type
Color Temperature
Lamp Life
Light Guide Connections
Weight in Pounds
Dimensions in Inches
Transport & Storage Conditions
Operating Conditions
Conditions of Visibility
4. CLASSIFICATIONS
Parameter
System Classification
Safety Certification
EMC Certifications
CE Marking
Protection against electric shock.
Degree of protection against harmful
ingress of water
Degree of safety in the presence of
flammable anesthetics
Mode of Operation
LIT-263 Sunoptic Technologies
Rev. A
Value
DPLS-01
Voltage: 100-240 Vac
Frequency: 50/60 Hz
Power: 450 Watts (maximum)
Ceramic 300W Xenon Lamp
Part No. SSXP300
5600K Nominal
1000 Hours
ACMI, Olympus, Storz, Wolf
15.8 lbs.
12.0 L x 14.0 W x 5.5 H
-4 to 140⁰F (-20 to 60⁰C)
0 to 95% RH (non-condensing)
70 to 106 kPa
50 to 86⁰F (10 to 30⁰C)
30 to 75% RH (non-condensing)
70 to 106 kPa
3 ft (1 m) distance, ±30⁰ viewing angle
Value
FDA Class I
EU Class I
Health Canada Class I
BF-type
IEC 60601-1 Ed. 3.1: 2012
CISPR 11 Class A, IEC 60601-1-2 4
Electrostatic discharge: ±8 kV contact, ±15 kV air
Radiated RF EM Fields: 3 V/m, 80 – 2700 MHz
EFT / Burst: ±2 kV, ±1 kV signal lines, 100 kHz
SURGE: ±0.5, ±1 kV
Conducted disturbance: 3 V 150 kHz – 80 MHz and 6 V in ISM
bands
Power frequency magnetic fields: 30 A/m
Voltage dips: 0%
225°, 270°, 315°, 40 %
Voltage interruptions: 0% for 5000 ms
Proximity fields: in accordance with EN 60601-1-2: 2015 table 9
Regulation (EU) 2017/745
Class I (grounded)
IPX-0 (no protection)
Device is NOT suitable for use in the presence of flammable
anesthetics.
Continuous
®
Page 6 of 134
th
Edition
/ 0.5 cycles at 0°, 45°, 90°, 135°, 180°,
/ 5 cycles, 70 %
/ 25 cycles
(English)