7. OPERATION
7.1 Turn ON
NOTE: Ensure that the unit is functional before attempting to use in a surgical procedure.
Connect power cord and insert fiber optic cable. Ensure lamp is seated and door is closed.
Flip rear AC power switch to ON (illuminates green) then press front panel ON / OFF button.
The front ON / OFF button illuminates blue when ON.
NOTE: the light source will default to 50% brightness upon initialization.
7.2 Functions
Light Setting
•
Press MINUS button to decrease light output by 10%.
•
Press and hold MINUS to decrease output by 10% until setting reaches 0%.
•
Press PLUS button to increase light output by 10%.
•
Press and hold PLUS to increase output by 10% until setting reaches 100%.
•
Light output setting is saved whenever changed, and the device will turn-on with the last
known light setting.
Standby
•
Press STANDBY button to temporarily block light output.
•
Light output decreases to 0%.
•
Standby LED illuminates and display indicates standby.
•
Press STANDBY button again to exit standby and return to previous brightness setting.
Menu
•
Press MENU button to view the following:
o
o
o
•
Lamp time reset: Press and hold STANDBY button until STANDBY LED illuminates (two
seconds). Lamp time changes to zero hours on menu screen.
•
Unit time cannot be reset.
LIT-263 Sunoptic Technologies
Rev. A
Figure 2. Standby Screen
Lamp time in hours.
Device time in hours.
Device software version.
®