MAINTENANCE AND USE MANUAL
Parameter
Mixing Time
Mixing Speed
Pre-mixing
Pre-vacuum time
Inversion interval
Fig. 9: Select the programme and press the dial (2). With
the Programme Name flashing against a black
background (d1), rotate the dial (2) to go to the "Mixing
time" parameter (d2); press the dial (2) to go to
modification of parameter (d2.1): rotate the dial to set
the value (d2.2) and press again to confirm (d2.3).
Parameters relative to mixing speed (d3), pre-mixing
(d4), pre-vacuum time (d5) and the rotation inversion
interval (d6) can be programmed in a way similar way to
the "Mixing time" example.
Fig. 10: When using the elite mix always select a
recipient of a suitable size and prepare the material to be
mixed. To obtain a good final result pre-mix by hand.
ELite mix
EN
Description
Range
from
Mixing Duration
00:00 - 05:00
Motor Speed
100 - 450
Time dedicated to
0 - 25
pre-mixing at 100
rpm
Time interval from
2 - 60
achievement of
constant in-cup
vacuum to start of
mixing
Sets time after which
2 - 30
the direction of
spatula rotation is
inverted
Unit of
Step
to
Measure
min:sec
1 sec
rpm
10
sec
5
sec
1
sec
1
2
3
Default
Notes
00:00
.
350
.
0
Can be activated or
deactivated (by setting
to 0)
0
Can be activated or
deactivated (by setting
to less than 2)
0
Inversion is
deactivated by setting to
less than 2 seconds
P06 Abcde 06
d1
time mixing 00:00
P06 Abcde 06
d2
00:00
time mixing
d2.1
time mixing 00:00
d2.2
time mixing 05:00
d2.3
time mixing
05:00
P06 Abcde 06
d3
rpm
350
P06 Abcde 06
d4
00
pre-mix time
P06 Abcde 06
d5
pre-vac time
00
P06 Abcde 06
d6
reverse time
00
9
Figure
10
Figure
121 of 264