Error messages
No.
Meaning and possible causes
01
Insufficient light at ZERO.Poss.cause:Duct contami-
nated or light path blocked; samplestrongly coloured?
03
Division by 0. Poss.cause:Duct contaminated or light
path blocked; samplestrongly coloured?
20
Too much light at detector. Poss.cause: Cap not
applied; measurement in bright sunlight
21
Too much light at detector. Poss.cause: Cap not
applied; measurement in bright sunlight
22
LowBat occurred during measurement. Change
battery, clean contacts
•
Maintenance
1. Keep the instrument in a clean condition.
2. No solvent, abrasive materials or brushes
should be used to clean the instrument. Clean
the sample chamber using a soft tissue.
3. Dry the instrument when not in use, store under
ambient environmental conditions.
4. Removebattery if instrument is being stored for
an extended period.
•
Trouble shooting
1. Useonly blister tablets with black printing.
2. Do not touch the tablets. This can cause
contamination and give pour results.
3. Closethe blue lid on top of the sample chamber tightly
before zeroing/reading.
4. Always ensurethat the sample chamber, the lid
and the stirring rod are thoroughly rinsed when
changing from one test to the other.
Warranty
The waterproof Scuball is warranted to be free from
manufacturing defects for two years.