IMPORTANT SAFEGUARDS – PLEASE MAKE THIS PRECAUTIONS LIST
WARNING –
1.
THIS DEVICE IS NOT INTENDED FOR LIFE SUPPORT. It may stop operating due to power interruption but
no hazards to patient.
2.
If oxygen is used with the device, the oxygen flow must be turned off when the device is not in use.
Explanation of the Warning: When the device is not in operation and the oxygen flow is left on,
oxygen delivered into the tubing may accumulate within the device's enclosure and create a risk
of fire.
3.
Oxygen supports combustion. Oxygen should not be used while smoking or in the presence of an
open flame.
4.
Always ensured the device is generating airflow before the oxygen supply is turned. Always turn off
the oxygen supply before stopping the airflow from the device.
5.
This device should not be used in the vicinity of a flammable anesthetic mixture in combination
with oxygen or air and nitrous oxide.
6.
The airflow for breathing generated by this device may be as much as 7°C (12.6° F) higher than the
room temperature. This device should not be used if the room temperature is warmer than 35°C
(95°F) to prevent the airflow temperature from exceeding 40° C (104° F) and causing
your airway.
7.
If this device overheats, it will stop operating and show "Error 002" on the display. After cooling
down to proper temperature, the device can restart again.
8.
This machine should be used only with masks (and connectors) recommended by the manufacturer,
or by your physician or respiratory therapist. A mask should not be used unless the CPAP machine is
turned on and operating properly. The vent holes associated with the mask should never be
blocked for proper exhaling purpose. If the vent hole is blocked, the CPAP machine will stop and
show message "Error 002", after cooling down, please re-connect the power cord to reset the
machine.
9.
At low CPAP pressure, some exhaled gas may remain in the mask and be re-breathed.
AVAILABLE TO ALL USERS and PROVIDERS