± 1 kV line(s) to
line(s)
Surge
± 2 kV line(s) to
IEC61000-4-5
earth
Voltage dips,
Voltage Dips:
short
i) 100% reduction for 0.5 period,
interruptions and
ii) 100% reduction for 1 period,
voltage variations
iii) 30% reduction for 25/30 period,
on power supply
Voltage Interruptions:
input lines
100% reduction for 250/300 period
IEC61000-4-11
Power frequency
(50/60Hz)
30 A/m
magnetic field
IEC61000-4-8
3 Vrms
0,15 MHz – 80 MHz
Conducted RF
6 Vrms in ISM bands
IEC 61000-4-6
between 0,15 MHz
and 80 MHz
80 % AM at 1 kHz
3 V/m 80 MHz to
2.7 GHz
80 % AM at 1 kHz
Radiated RF EM
385-6000 MHz,
Fields
9-28V/m, 80%
IEC61000-4-3
AM(1kHz) pulse
mode and other
modulation
± 1 kV line(s) to
± 1 kV line(s) to
line(s)
line(s)
100- 240 V
30 A/m
30 A/m
3 Vrms
0,15 MHz – 80
MHz
6 Vrms in ISM and
amateur radio
6Vrms
bands between
0,15 MHz and 80
MHz
80 % AM at 1 kHz
10 V/m 80 MHz to
2,7 GHz
80 % AM at 1 kHz
385-6000 MHz,
10V/m
9-28V/m, 80%
AM(1kHz) pulse
mode and other
modulation
Mains power quality should
be that of atypical
commercial or hospital
environment.
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
this device requires
continued operation during
power mains interruptions,
it is recommended that the
device be powered from an
uninterruptible power
supply or a battery.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
Portable and mobile RF
communications equipment
should be used no closer to
any part of this device,
including cables, than there
commended separation
distance calculated from
the equation applicable to
the frequency of the
transmitter.
Recommended separation
distance
d
=
150kHz to 80MHz
d
=0.6
80MHz to 800MHz
d
=1.2
800 MHz to 2.7
GHz
Where P is the maximum
output power rating of the
transmitter in watts (W)
according to the transmitter
manufacturer and d is the
recommended separation
distance in meters (m).
Field strengths from fixed
RF transmitters, as
determined by an
electromagnetic site
a
survey ,
should be less
than the compliance level in
each frequency ranged.
Interference may occur in
the vicinity of equipment
marked with the following
symbol: