Steri-Mate, Steri-Mate 360 & Jet-Mate Handpieces, Handpiece Adapters, and Reservoir Bottles
ITEMS
• Jet-Mate Handpiece
• Black Steri-Mate Handpiece
• Handpiece Adapters
• Reservoir Bottles
• Blue Interface Caps with Sip Tubes
Gray Steri-Mate Handpiece
Steri-Mate 360° Handpiece
Alternate method: Place non-bagged instruments into the steam autoclave. After warm-up is completed, operate at the
sterilizing temperature and cycle time listed in the chart above. Instruments should be used immediately.
Drying:
To dry, use the drying cycle of the autoclave. Set cycle for 20-30 minutes.
Maintenance, inspection and testing:
• Visually inspect to ensure all contamination is removed.
• Check for distortion, damage and wear. Discard damaged, worn or corroded instruments.
• Inspect o-rings on the handpiece, handpiece adapter and handpiece cable for damage or wear. Replace if necessary.
• Ensure all o-rings are in place.
INFECTION CONTROL PROCEDURES
STERILIZING TEMPERATURE
270°F / 132°C
273°F / 134°C
270°F / 132°C
STERILIZING CYCLE TIME
3 minutes
4 minutes
10 minutes
7