INFECTION CONTROL PROCEDURES
Steri-Mate, Steri-Mate 360 & Jet-Mate Handpieces, Handpiece Adapters, and Reservoir Bottles
INSTRUCTIONS:
Point of Use:
Before processing, remove Ultrasonic insert from the handpiece and disconnect handpiece from the scaler cable.
Remove any moist debris/residue on the handpiece using a lint free wipe (disposable cloth or paper towel) moistened
with water. Then dry handpiece with a fresh dry paper towel or disposable cloth. Alternatively, use a disposable disinfec-
tant cleaning wipe (such as VoloWipes™) to remove moist debris/residue.
Do not allow residue or any form of contamination to dry on device.
Containment and Transportation:
Protect instruments from damage. Instruments should be reprocessed as soon as is reasonably practical following use.
Instruments left standing wet may stain or corrode.
Cleaning: manual
Follow manufacturer instructions for cleaning detergent observing concentration rates and contact times.To remove con-
tamination thoroughly wipe the handpiece using soap and water or a cleaning solution (such as ReSURGE
solution);
®
pay special attention to holes/cannulas – these can be cleaned using a disposable syringe filled with soap and water or
cleaning solution (such as ReSURGE
).
®
Rinsing:
Rinse the handpiece under tepid (potable) running water for 30 seconds to remove cleaning agent; a disposable syringe
may be used to rinse the holes/cannulas.