INFECTION CONTROL PROCEDURES
Cavitron Systems
Containment and Transportation:
Handle with care.
Cleaning: manual
Generously spray disinfectant solution on a clean cloth. Wipe the outer surfaces of the cabinet, power cords, non-steril-
izable handpieces, handpiece cables, air and water lines, foot control, foot control cables, and infuser bag. Discard used
cloth. Wipe dry with a clean cloth.
Disinfection:
Generously spray disinfectant solution on a clean cloth. Wipe the outer surfaces of the cabinet, power cords, non-steril-
izable handpieces, handpiece cables, air and water lines, foot control, foot control cables, and infuser bag. Discard used
cloth. Allow disinfectant to air dry.
Drying:
When cleaning, wipe surfaces dry with a clean cloth. To achieve disinfection, allow surfaces to air dry.
Maintenance, inspection and testing:
• Visually inspect to ensure all contamination is removed.
• Visually inspect all cables and power cord for tears.
• Visually inspect handpiece cable o-rings for wear. Replace worn o-rings (refer to applicable Directions for Use of sys-
tem). Lubricate o-rings with water before use.