TABLE 8: Scan head dimensions
Dimension
Width
Depth
Height
Weight
Door access width
4.1.4 Review client properties
Browser-based review clients let you review and analyze patient images.
TABLE 9: Review client properties
Review clients
Display properties for browser-based review clients.
The minimum recommended monitor resolution.
Recommended browsers for OptosAdvance.
4.1.5 Touch screen properties
The touch screen is powered from the scan head at +12V.
Do
NOT
WARNING
TABLE 10: Touch screen properties
Dimension
Minimum display resolution
Width
Depth
Height
Weight
Touch screen mount
4.1.6 Table requirements
Do
NOT
plugged directly into a wall socket.
The scan head should be positioned in the middle of the table width. The front of the scan
head should be inline with the front of the table to allow the patient to get into the correct
WARNING
alignment position.
The chin rest should slightly overhang the edge of the table.
Part Number: G108025/4GCA
Copyright 2017, Optos plc. All rights reserved.
Details
550 mm (22 inches)
550 mm (22 inches) including chin rest
608 mm to 632 mm (24 inches to 25 inches)
34 kg (75 lbs)
760 mm (30 inches)
disconnect or connect the touch screen while the scan head is in use.
Details
1024 x 768 pixels, 16.7 million colors
260 mm (10.5 inches)
20 mm (1 inch)
216 mm (8.5 inches)
less than 1 kg
VESA MIS-D, 75, C (hole pattern: 75 mm x 75 mm)
plug the scan head into a power block or powered table. The scan head must be
Chapter 4 - Technical Specification
1280 x 1024 pixels, 16.7 million colors.
Google Chrome (46.0 and higher)
Internet Explorer (11.0 and higher)
Apple Safari (6.0 and higher)
Page 57 of 69
English