GUIDANCE AND MANUFACTURER'S DECLARATION
The FarmaDolor. is intended for use in the electromagnetic environment specified below.
The customer or the user of the FarmaDolor. should assure that it is used in such an
IEC 60601
IMMUNITY TEST
TEST LEVEL
3 Vrms
Conducted RF IEC
61000-4-6
150 kHz to 80 MHz
3 V/m
Radiated RF IEC
61000-4-3
80 MHz to 2.5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment
due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the FarmaDolor is used exceeds the applicable
RF compliance level above, the FarmaDolor should be observed to verify normal operation. If
abnormal performance is observed, additional measures may be necessary, such as re-orienting
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
ELECTROMAGNETIC IMMUNITY
environment.
COMPLIANCE
LEVEL
equipment should be used no closer to any part
of the FarmaDolor, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
Not applicable
where P is the maximum output power rating
recommended separation Distance in metres (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey,a
should be less than the compliance level in each
3 V/m
equipment marked with the following symbol:
or relocating the FarmaDolor.
ELECTROMAGNETIC ENVIRONMENT –
GUIDANCE
Portable and mobile RF communications
the transmitter.
Recommended separation distance
d = 1.2
d = 1.2
80 MHz to 800 MHz
d = 2.3
80 MHz to 2.5 GHz
of the transmitter in watts (W) according to
the transmitter manufacturer and d is the
frequency range.b
Interference may occur in the vicinity of